EasyManua.ls Logo

Fluke 724 - Millivolt Measurement and Source Specifications; Temperature and Thermocouple Specifications

Fluke 724
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Temperature Calibrator
Specifications
41
Millivolt Measurement and Source*
Range Resolution Accuracy
-10 mV to 75 mV 0.01 mV ±(0.025 % + 1 count)
Maximum input voltage: 30 V
Temperature coefficient -10 °C to 18 °C, +28 °C to
55 °C: ±0.005 % of range per °C
*Select this function by pressing T. The signal is
available at the thermocouple miniplug connector.
Temperature, Thermocouples
Type Range
Measure and Source
Accuracies (ITS-90)
J -200 to 0 °C
0 to 1200 °C
1.0 °C
0.7 °C
K -200 to 0 °C
0 to 1370 °C
1.2 °C
0.8 °C
T -200 to 0 °C
0 to 400 °C
1.2 °C
0.8 °C
E -200 to 0 °C
0 to 950 °C
0.9 °C
0.7 °C
R -20 to 0 °C
0 to 500 °C
500 to 1750 °C
2.5 °C
1.8 °C
1.4 °C
S -20 to 0 °C
0 to 500 °C
500 to 1750 °C
2.5 °C
1.8 °C
1.5 °C
B 600 to 800 °C
800 to 1000 °C
1000 to 1800 °C
2.2 °C
1.8 °C
1.4 °C
L -200 to 0 °C
0 to 900 °C
0.85 °C
0.7 °C
U -200 to 0 °C
0 to 400 °C
1.1 °C
0.75 °C
N -200 to 0 °C
0 to 1300 °C
1.5 °C
0.9 °C
Resolution:
J, K, T, E, L, N, U: 0.1 °C, 0.1 °F
B, R, S: 1 °C, 1 °F

Other manuals for Fluke 724