EasyManua.ls Logo

Fluke 724 - Ohms Measurement Specifications; Ohms Source Specifications

Fluke 724
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
724
Users Manual
40
Ohms Measurement
Accuracy ±
Ohms Range 4-Wire 2- and 3-Wire*
0 to 400 0.1 0.15
400 to 1.5 k 0.5 1.0
1.5 to 3.2 k 11.5
Excitation Current: 0.2 mA
Maximum input voltage: 30 V
Temperature coefficient -10 °C to 18 °C, +28 °C to
55 °C: ± 0.005 % of range per °C
* 2-wire: Does not include lead resistance.
3-wire: Assumes matched leads with a total resistance
not exceeding 100 .
Ohms Source
Ohms Range
Excitation Current from
Measurement Device
Accuracy
±
15 to 400 0.15 to 0.5 mA 0.15
15 to 400 0.5 to 2 mA 0.1
400 to 1.5 k 0.05 to 0.8 mA 0.5
1.5 to 3.2 k 0.05 to 0.4 mA 1
Resolution
15 to 400 0.1
400 to 3.2 k 1
Temperature coefficient -10 °C to 18 °C, +28 °C to
55 °C: ± 0.005 % of resistance range per °C