724
Users Manual
40
Ohms Measurement
Accuracy ±
Ohms Range 4-Wire 2- and 3-Wire*
0 to 400 0.1 0.15
400 to 1.5 k 0.5 1.0
1.5 to 3.2 k 11.5
Excitation Current: 0.2 mA
Maximum input voltage: 30 V
Temperature coefficient -10 °C to 18 °C, +28 °C to
55 °C: ± 0.005 % of range per °C
* 2-wire: Does not include lead resistance.
3-wire: Assumes matched leads with a total resistance
not exceeding 100 .
Ohms Source
Ohms Range
Excitation Current from
Measurement Device
Accuracy
±
15 to 400 0.15 to 0.5 mA 0.15
15 to 400 0.5 to 2 mA 0.1
400 to 1.5 k 0.05 to 0.8 mA 0.5
1.5 to 3.2 k 0.05 to 0.4 mA 1
Resolution
15 to 400 0.1
400 to 3.2 k 1
Temperature coefficient -10 °C to 18 °C, +28 °C to
55 °C: ± 0.005 % of resistance range per °C

Other manuals for Fluke 724

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 724 and is the answer not in the manual?

Fluke 724 Specifications

General IconGeneral
Temperature Range-10°C to 55°C
Thermocouple TypesJ, K, T, E, R, S, B, L, U, N
Current Measurement0 to 24 mA
Operating Temperature-10°C to 55°C
FunctionProcess calibrator
RTD TypesPt100, Pt500, Pt1000
Voltage Measurement0 to 10 V
Loop Power24 V DC
DisplayDual Display
Power Supply4 AA batteries
Resolution0.1 °C
Storage Temperature-20 °C to 60 °C
Battery Life25 hours
Relative Humidity0 % to 90 %
Safety ComplianceCSA C22.2 No. 1010.1:1992