● Do not use the control solution beyond the expiry
date or 3 months after rst opening. Write the
opening date on the control solution vial and
discard the remaining solution after 3 months.
● It is recommended that the control solution test
be done at room temperature 20°C to 25°C (68°F
to77°F). Make sure your control solution, meter,
and test strips are at this specied temperature
range before testing.
● Shake the vial before use, discard the rst drop of
control solution, and wipe off the dispenser tip to
ensure a pure sample and an accurate result.
● Store the control solution tightly closed at
temperatures between 2°C to 30°C (35.6°F to
86°F). Do NOT freeze.
FORA iDiamond Meter
Specications
Model No.: DM40
Dimension & Weight: 55 (L) x 36 (W) x 12.10 (H)
mm, 13.6 g
Power Source: iOS device.
Memory: Control by iOS device.
External Output: Apple 30 pins Connector
Auto electrode insertion detection
Auto sample loading detection
Temperature warning
Operating Condition: 10°C to 40°C (50°F to 104°F),
below 85% R.H. (non-condensing)
Meter Storage/ Transportation Conditions: -20°C to
60°C (-4°F to 140°F), below 95% R.H.
Strip Storage/Transport Conditions: 2°C to 32°C
(35.6°F to 89.6°F ), below 85% R.H.
Measurement Units: Control by iFORA
Measurement Range: 20 to 600mg/dL (1.1 to
33.3mmol/L)
This device has been tested to meet the electrical
and safety requirements of: IEC/EN 61010-1, IEC/
EN 61010-2-101, EN 61326-1, IEC/EN 61326-2-6,
EN 301 489-17, EN 300 328.