Cleaning
1. To clean the meter exterior, wipe it with a cloth
moistened with tap water or a mild cleaning
agent, then dry the device with a soft dry cloth.
Do NOT rinse with water.
2. Do NOT use organic solvents to clean the meter.
Meter Storage
● Storage conditions: -20°C to 60°C (-4°F to140°F),
below 95% relative humidity.
● Always store or transport the meter in its original
storage case.
● Avoid dropping and heavy impact.
● Avoid direct sunlight and high humidity.
Meter Disposal
The used meter should be treated as contaminated
that may carry a risk of infection during
measurement. The batteries in this used meter
should be removed and the meter should be
disposed in accordance with local regulations.
The meter falls outside the scope of the European
Directive 2002/96/EC-Directive on waste electrical
and electronic equipment (WEEE).
Caring for Your Test Strips
● Storage conditions: 2°C to 32°C (35.6°F to
89.6°F), below 85% relative humidity. Do not
freeze.
● Store your test strips in their original vial only. Do
not transfer to another container.
● Store test strip packages in a cool dry place.
Keep away from direct sunlight and heat.
● After removing a test strip from the vial,
immediately close the vial cap tightly.
● Touch the test strip with clean and dry hands.
● Use each test strip immediately after removing it
from the vial.
● Write the opening date on the vial label when you
rst opened it. Discard remaining test strips after
3 months.
● Do not use test strips beyond the expiry date.
This may cause inaccurate results.
● Do not bend, cut, or alter a test strip in any way.
● Keep the strip vial away from children since the
cap and the test strip may be a choking hazard. If
swallowed, promptly see a doctor for help.
For further information, please refer to the test strip
package insert.
Important Control Solution Information
● Use only FORA control solutions with your meter.