When “QC” is selected, the control solution test
result will appear on the screen however the meter
will not store the test result in the memory of the
iOS device.
WARNING:
When doing the control solution test, you have to
mark it so that the test result will not be stored in
the memory. Failure to do so will mix up the blood
glucose test results with the control solution test
results in memory.
3. Apply Control Solution
Shake the control solution vial thoroughly before
use.
Squeeze out a drop and wipe it off, then squeeze
out another drop and place it on the tip of the
vial cap. Hold the meter to move the absorbent
hole of the test strip to touch the drop. Once the
conrmation window lls completely, the meter will
start processing. To avoid contaminating the control
solution, do not directly apply control solution onto a
strip.