EasyManua.ls Logo

Freestyle LIBRE PRO - Page 23

Freestyle LIBRE PRO
49 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Reader Specications
Reader size 95 mm x 60 mm x 16 mm
Reader weight 65 grams
Reader power source One lithium-ion rechargeable battery
Reader battery life 2 weeks of typical use
Reader Sensor memory 1 Sensor
Reader operating
temperature
50 °F to 113 °F
Reader storage
temperature
-4 °F to 140 °F
Operating and storage
relative humidity
10-90%, non-condensing
Reader moisture
protection
Keep dry
Operating and storage
altitude
-1,250 ft (-381 meters) to 10,000 ft
(3,048 meters)
Reader display timeout 60 seconds
Radio Frequency
Near Field Communication*
(13.56 MHz RFID); ASK Modulation;
124 dBuV/m;
1.5 inch communication range
Data port Micro USB
Minimum Computer
Requirements
System must only be used with
EN60950-1 rated computers
Mean service life 3 years of typical use
Power Adapter
Abbott Diabetes Care PRT25611
Operating temperature: 50 °F to 104 °F
USB Cable
Abbott Diabetes Care PRT21373
Length: 37 inches (94 cm)
* Security measures: The communication between Reader and Sensor is a short range near
eld communication method making it dicult to interfere with or intercept data that is
being transferred. The Sensor and Reader are protected by proprietary data format, memory
mapping, and cyclic redundancy check (CRC) generation and verication of data.
Quality of Service (QoS): QoS for the FreeStyle Libre Pro Reader and Sensor wireless
communications using the near eld communications is assured within the eective range
of 4 cm between the Sensor and Reader that is specied to occur within 15 seconds.
40 41
ART26944_rev-C_manual.indd 40-41 9/23/16 9:19 AM

Related product manuals