EasyManua.ls Logo

FTS Bubbler - Page 39

FTS Bubbler
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
700-Bubbler-Man Rev 12 13 Feb 2020
Page 34/51
COMMAND(S)
RESULTS
aM!
aC!
aMC!
aCC!
Measures:
Stage
Units of stage*
Status *
The status bits will only be reset if the data logger performs an XCON
command or XDT for the SW_SYNC_TIME.
Some HW bits can be reset.
aM1!
aC1!
aMC1!
aCC1!
Measures:
Stage
Units of stage*
Temperature
Units of Temperature*
Offset (Units same as Stage)
Low Accuracy Line Pressure (psi)
Tank Pressure (psi)
Battery Voltage (volts)
Status *
This is not the same line pressure source which is used to calculate stage. Stage measurements
are determined using a high accuracy transducer. The low accuracy transducer should only be
used as a troubleshooting tool as it provides insight into the pressure status of the system.
aM2!
aC2!
aMC2!
aCC2!
Measures:
Purge state*
Pump on time
Pump on-off count
Pump fault count
Uncompensated stage
Status *
aM7!
aC7!
aMC7!
aCC7!
Measures:
Purge state*
Purge pressure
Tank pressure
Status*
See section 5.1.2 for details of purge actions using the M7 command.
Use XPR to set purge pressure (default = 90 psi)
*See Section 5.1.1 for measurement codes and meanings

Table of Contents

Other manuals for FTS Bubbler

Related product manuals