700-Bubbler-Man Rev 12 13 Feb 2020
Page 34/51
COMMAND(S)
RESULTS
aM!
aC!
aMC!
aCC!
Measures:
• Stage
• Units of stage*
• Status *
The status bits will only be reset if the data logger performs an XCON
command or XDT for the SW_SYNC_TIME.
Some HW bits can be reset.
aM1!
aC1!
aMC1!
aCC1!
Measures:
• Stage
• Units of stage*
• Temperature
• Units of Temperature*
• Offset (Units same as Stage)
• Low Accuracy Line Pressure (psi)
†
• Tank Pressure (psi)
• Battery Voltage (volts)
• Status *
† This is not the same line pressure source which is used to calculate stage. Stage measurements
are determined using a high accuracy transducer. The low accuracy transducer should only be
used as a troubleshooting tool as it provides insight into the pressure status of the system.
aM2!
aC2!
aMC2!
aCC2!
Measures:
• Purge state*
• Pump on time
• Pump on-off count
• Pump fault count
• Uncompensated stage
•
Status *
aM7!
aC7!
aMC7!
aCC7!
Measures:
• Purge state*
• Purge pressure
• Tank pressure
• Status*
See section 5.1.2 for details of purge actions using the M7 command.
Use XPR to set purge pressure (default = 90 psi)
*See Section 5.1.1 for measurement codes and meanings