5.2.3 X Commands List
Begin tank bleed If
is empty or
=B
Bubbler needs to be power-cycled after this command
in order to return to normal operations.
aXBR!
aXBRP!
Get bubble rate
Set bubble rate
Bubbles per minute. Minimum value is 30 and max
value is 120, default is 60.
To set bubble rate, replace P with desired bubble rate
aXBVI!
aXBVIP.p!
Get battery voltage
idle
Set battery voltage
idle
P.p is the minimum voltage that will trigger a voltage
fault,
Default = 10.8.
aXBVL!
aXBVLP.p!
Get battery voltage
low
Set battery voltage
low
P.p is the minimum voltage that will trigger a battery
fault,
Default = 8.0
aXBVH!
a XBVHP.p!
Get battery voltage
high
Set battery voltage
high
P.p is the maximum voltage that will trigger a battery
fault.
Default = 16.5
aXCON!
change
Reset the configuration change flag indicating the
settings are in sync with the data logger.
aXDOC!
Get date of
calibration
Returns in YYYYMMDD format.
aXDOM!
manufacture
Returns in YYYYMMDD format.
Returns date and time in YYYYMMDD,HHMMSS format.
aXDTYYYYMMDD,
HHMMSS!
Set date and time
Example:
a20170125,130000! Sets the Bubbler’s date
to 2017 January 25 1 p.m.
Date range is from 20170101 – 21050101 (Jan 1, 2017
to Jan 1, 2105)
Time range is from 000000 to 235959