700-Bubbler-Man Rev 12 13 Feb 2020
Page 38/51
X Command
Definition
Comments
aXFSN!
Get serial number
Example: 123456
aXI!
Get firmware
version and build
date
Example: 1.0.65,2017-06-21 17:19:29 UTC
aXID!
aXIDP!
Get site ID
Set site ID
Site ID is an alphanumeric string of maximum 10
characters.
aXID!abubbler1 (gets Site ID of “bubbler1”)
aXIDbubbler1! (sets Site ID to bubbler 1)
There is no error checking for symbols. Symbols can be
entered but are not supported on the Bubbler’s
display.
aXIMIN!
aXIMINP!
Get 4-20 mA
minimum
Set 4-20 mA
minimum
Returns the set minimum value
Replace P with desired value to set minimum
Default = 0.0
The minimum and maximum values must differ by
0.2 psi (6 inches). Stored in psi
aXIMAX!
aXIMAXP!
Get 4-20 mA
maximum
Set 4-20 mA
maximum
Returns the set maximum value
Replace P with desired value to set maximum value
Default = 0.0
The minimum and maximum values must differ by
0.2 psi (6 inches). Stored in psi
aXPCB!
Get hardware
revision
This is derived by a resistor set in the hardware.
aXPUMP!
aXPUMPP!
Get pump
(compressor)
statistics
Set pump
(compressor)
statistics
Returns on-time, on-off count, and fault count.
P=Z then statistics are zeroed
aXPR_!
Get purge pressure
Return purge pressure setting in psi
aXPR_P_!
Begin/end purge
or
If P is B = begin purge mode
If P is E = end purge mode.
Set purge pressure
Replace P with desired purge pressure (in psi)
Example: aXPR90! (set purge pressure at 90 psi)
Valid range is 20 to 90. Default :90