5
If there is a danger of a serious accident resulting from a failure or a defect in this unit, provide the
unit with an appropriate external protective circuit to prevent an accident.
The unit is supplied without a power switch and fuses.
Make wiring so that the fuse is placed between the main power supply switch and this controller.
(Main power supply: 2 pole breaker, fuse rating: 250V, 1A)
When wiring the power supply terminal, use vinyl insulated 600 volt cable or equivalent.
To avoid the damage and failure of controller, supply the power voltage tting to the rating.
To avoid an electric shock and controller failure, do not turn ON the power before all wiring is
completed.
Be sure to check that the distance is kept to avoid electric shock or ring before turning the power
ON.
Keep away from terminals while the circuit is energized in order to avoid an electric shock and a
malfunction.
Never attempt to disassemble, fabricate, modify, or repair this unit because tampering with the unit
may result in a malfunction, electric shock, or a re.
Output relay is the part has a limited life.
When output relay contact comes to the end of its life, it might remain on-
state, or off-state. For safety, use a protective circuit outside.
1.2 Maintenance precautions
Be sure to turn off the power before this controller is installed or removed in order to avoid an
electric shock, malfunction, and fault.
Regular maintenance is recommended a longer service life of this controller. Some parts of this
controller have a limited life span, or they will be deteriorated with the lapse of time.
One-year warranty is guaranteed for this unit including accessories, prvided that the controller is
properly used.

Other manuals for Fuji Electric PXR3