Preparation before use
8
Installation of the main unit
Place the chair at a level place by securing sufficient
spaces around it.
Request: As the back rest is reclined, the foot rest rises
simultaneously. Minimum 60 cm of spaces must be
secured both at the front and the back of the chair.
Request: Bottom of the chair may damage the tatami
mattress or floor. It is recommended to spread a
mattress, etc. under the chair.
Changing the place of installation
Raising the front of main unit, push it to move.
CAUTION
Do not attempt to move the chair while a person or object
sitting on the seat. There is risk of upset.
Where the floor surface is not hard enough, the chair should
not be moved using the casters or by dragging directly on the
floor.
Do not hold the seat, foot rest or shoulder section to move the chair.
Front part of the chair is rather heavy. Take sufficient care when
lifting the front part of the chair..
Grounding connection
Securely connect the grounding cable at
the chair.
Unless the grounding cable is
installed, it could cause electrical
shocks when the earth leakage
occurs. Consult your dealer to
install the grounding cable.
When the grounding terminal is provided
at the wall receptacle.
Extend the grounding cable
from the grounding terminal
on the chair main unit to that
on the receptacle.
Gas pipe … There is risk of explosion or
ignition.
Telephone cable or lightning rod … It is
dangerous when struck by lightning.
Water pipe … Earth connection is not
established if plastic pipes are connected
on the way.
When the grounding terminal is not
provided at the wall receptacle.
Consult your dealer or electrical work
contractor for the installation of grounding
cable.
CAUTION
Do not connect the grounding cable
at the followings:
Grounding cable
60 cm or more
60 cm or more
Casters
*
(Grounding terminals are provided on the models for
Thailand, Sweden, France and Russia.
.