4-1
Chapter 4 Entering Patient Information
Enter the name, age and gender of the patient.
There are following 6 methods to enter the information.
Select an appropriate method depending on the situation.
The Difference between Patient File Selection and Loading Past Information
• Patient file selection allows to select patient information from a patient information list.
• Loading past information allows to load patient information from examination data saved to a media (including internal
memory) by entering the patient's ID. The patient's height can be loaded from previous examination data, which is convenient
when measuring CAVI in pulse wave examinations. (This is the same as the "Load ID" function of the VS-1500 series.)
When using the device for the first time, set whether to assign the ID in the patient information with or without zero
padding (adding zeros to the ID). If using zero padding, set the number of digits to be used for ID. (Refer to "Patient
Information Settings" (P11-9).)
Patient Data Categories
The following information can be entered.
Input Method Description Reference
Basic
information
input
Enter items such as the ID, age, gender and height
manually.
"Entering the Basic Information" (P4-3)
Master ID Load information from the DMS. "Loading Information from the DMS (Master ID)"
(P4-8)
Patient file
selection
Use patient information registered to an SD card. "Using Patient Information Registered to an SD
Card (Patient File Selection)" (P4-9)
Load past
information
Use patient information from examination data saved to
the media (including internal memory).
"Using Patient Information from Examination
Data Saved to the Media (Load Past Information)"
(P4-12)
External input Read information from a magnetic card, barcode, etc. "Connecting an ID Reader and Displaying Order
Information" (P10-8)
Display Description Setting Range Note
ID Patient code 3-20 digits
Can be entered in alphanumeric characters
Age Age 3 digits (0-150) Separate input fields are used for the
year, month, week and day. (Do not
enter months, weeks or days for
CAVI measurements.)
A date of birth can also be entered.
Gender Gender Male/Female
Name Name 24 single-byte characters Alphanumeric characters and
symbols can be entered.
Dept. Department information such as the
ward (Dept. 1) and department
(Dept. 2)
Up to 4 items can be entered in
Dept. 1-4.
Code: 4-8 digits
16 single-byte characters
4-8 digits, can be changed
A total of up to 400 items can be
registered for Dept. 1-4.
Height Height 3-digit integer + 1 decimal place Unit: cm, inch
Max. 300cm
Weight Weight 3-digit integer + 1 decimal place Unit: kg, lb
Max. 300kg

Table of Contents