Calibration
32
Potentiostat calibration is only required infrequently. Recalibrate your Interface 5000 under the following
circumstances:
It is at least one year since your last calibration.
Your potentiostat has been serviced.
You notice breaks or discontinuities in the data curves recorded with your system.
The system is being run in an environment that is very different from the previous operating environment.
For example, if the Interface 5000 was calibrated at 15°C and you are now operating it at 30°C, you should
recalibrate.
Figure 4-1
Interface 5000 Calibration Cell with Leads Attached for Calibration
Cable Calibration
This procedure uses an external resistive Calibration Cell, the same one as for DC Calibration: