110
O
2
measurement
NOTE: B40 only.
Selecting the O
2
scale
If the difference between FiO
2
and EtO
2
is above 6%, change the O
2
scale:
1. Press the
Airway Gas key.
2. Select O2 Setup.
3. Select Scale
4. Select an option from the scale list.
OFF/ON the O
2
measurement
1. Press the Airway Gas key.
2. Select O2 Setup - Measurement.
3. Choose OFF or ON
Setting O
2
alarms
1. Press the
Airway Gas key.
2. Select O2 Setup - O2 Alarm.
3. Select Adjust Limits to set up the limits.
4. Select EtO2 Alarm, FiO2 Alarm to OFF/ON the limits.
5. You can also select Alarms Setup to adjust O
2
or related alarms.
Anesthetic agent and N
2
O measurement
NOTE: B40 only.
NOTE: The E-sCAiO and N-CAiO modules automatically identify the
agent being used. If the agent value is out of measurment range,
ā€œ---ā€ will be displayed on screen, which means the value is invalid.
Selecting the agent scale
1. Press the Airway Gas key.
2. Select Agent/N2O Setup - Agent Scale
3. Select an option from the scale list.
OFF/ON the agent measurement
1. Press the Airway Gas key.
2. Select Agent/N2O Setup - Agent Measurement.
3. Choose OFF or ON
OFF/ON the N
2
O measurement
1. Press the Airway Gas key.
2. Select Agent/N2O Setup - N2O Measurement.
3. Choose OFF or ON