EasyManua.ls Logo

GE B20 - Page 126

GE B20
160 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
122
Entropy
Entropy values seem unstable:
Check that the sensor is not dried out.
Check the sensor attachment and placement.
Check the patient status.
Entropy EEG signal is noisy:
Remove disturbing equipment from the proximity of the
Entropy module or sensor.
Check the sensor’s contact with skin.
Check electrodes.
Entropy EEG signal is poor:
Check the sensor’s contact with skin.
Check electrodes.
Entropy waveform and numbers do not correspond:
Check raw EEG as impedance check may cause a temporary
increase in the numeric values.
Check the patient’s overall status.
Entropy readings seem inconsistent with the patient status:
Check raw EEG for QRS or other artifact.
Check electrode placement.
Impedance respiration
Measurement fails:
Check the electrode quality and positioning.
Adjust the detection limits. During ventilator supported
breathing, the respiration calculation may count only
ventilator-produced inspirations and expirations.
Other electrical devices may interfere with the
measurement.
Invasive pressures
Readings seem unstable:
Make sure that there are no air bubbles in the transducer
system.
Flush and zero.
Place the transducer on the patient’s mid-heart level and
zero.
Invasive blood pressure waveform is displayed but no numeric
values are displayed:
Zero the channel. Invasive blood pressure numerical values
are displayed only for successfully zeroed channels.
Monitor
The monitor does not start:
Check that the batteries are inserted and sufficiently
charged, see page 25.
Check that the power cord is properly connected.
Check the fuses and replace them if necessary, see “Cleaning
and care.”
Measurements
The measured values are not displayed:
Check that you have selected the desired parameter to a
waveform or digit field, seeModifying the screen setup.”
You cannot perform a measurement or a function.
Check that the measurement module is properly installed.
Remove the module and reinstall it.

Related product manuals