Table 2: The table below shows Arms values measured using GE SpO
2
sensors with GE CARESCAPE
TM
V100 in a clinical study.*
* The sensors were clinically tested for accuracy with the following sensors: OXY-E (equivalent to TS-E-D, TS-E2-GE, TS-E4-GE), OXY-SE (equivalent to TS-SE-3),
OXY-F (equivalent to TS-F-D, TS-F2-GE, TS-F4-GE, TS-SA-D, TS-SA4-GE), OXY-W (equivalent to TS-W-D), OXY-AP (equivalent to TS-AP-10, TS-AP-25), OXY-AF
(equivalent to TS-AF-10, TS-AF-25).
The sensors were clinically tested for neonatal accuracy with the following sensors: OXY-SE (equivalent to TS-SE-3), OXY-AF (equivalent to TS-AF-10, TS-AF-25).
For supplemental data analysis presented in a format of Bland-Altman in User’s Reference Manual 2081504-001, please check the
accompanying document or CD.
GE SpO
2
Sensor 70 - 80% 80 - 90% 90 - 100%
OXY-E 2.3 digits 1.4 digits 1.3 digits
OXY-SE 2.5 digits 2.0 digits 1.1 digits
OXY-F 1.3 digits 1.0 digits 1.1 digits
OXY-W 2.9 digits 1.8 digits 1.0 digits
OXY-AP 2.0 digits 1.9 digits 1.7 digits
OXY-AF 2.5 digits 1.4 digits 0.9 digits