GE B20 - Airway gases (continued); Anesthetic agents measurement specifications

GE B20
160 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
142
Nellcor SpO
2
specifications
Masimo SpO
2
specifications
Measurement range 1 to 100%
Display range 0 to 100%
Calibrated against functional oxygen saturation.
Measurement accuracy Adult 100 to 70% ±2 digits
Neo 100 to 70% ±3 digits
Low perfusion 100 to 70% ±2 digits
Display resolution 1% of SpO
2
Display averaging 2 to 7 seconds
Pulse rate
Measurement and display
range
20 to 250 bpm
Display resolution 1 bpm
Measurement accuracy ±3 digits
Sensor Light Source*
Wavelength Infrared: 900 nm
Red: 660 nm
Total optical output power
of the sensor LEDs
less than 15 mW
* This information may be useful to clinicians, such as those performing
photodynamic therapy.
Measurement range 1 to 100%
Display range 0 to 100%
Calibrated against functional oxygen saturation.
Measurement accuracy
Without motion Adult/Pediatric 100 to 70% ±2 digits
Neonate 100 to 70% ±3 digits
With motion Adult/Ped/Neo 100 to 70% ±3 digits
Low perfusion 100 to 70% ±2 digits
0~69% unspecified
Display resolution 1% of SpO
2
Display averaging 2 to 16 seconds
Pulse rate
Measurement and display
range
25 to 240 bpm
Display resolution 1 bpm
Measurement accuracy Without motion ±3 bpm
With motion ±5 bpm
Sensor Light Source*
Wavelength Infrared: 905 nm (nominal)
Red: 660 nm (nominal)
Power Dissipation Infrared: 22.5 mW (max)
Red: 27.5 mW (max)