143
Invasive blood pressure
NIBP
* This information may be useful to clinicians, such as those performing
photodynamic therapy.
Measurement range -40 to 320 mmHg (-5.3 to 42.7 kPa)
Measurement accuracy ±5% or ±2 mmHg, whichever is
greater
Frequency response 4 to 22 Hz
Transducer sensitivity 5 µV/V/mmHg
Pulse rate
Range 30 to 250 bpm
Accuracy ±5% or ±5 bpm, whichever is greater
Display resolution 1 bpm
Zero adjustment range ± 150 mmHg
Measurement technique Oscillometric with step deflation
Supported modes Manual, automatic and stat
Measurement time Adult/Pediatric inflate duration time
less than 120 s
Neonate cycle time less than 85 s
Measurement ranges
Systolic Adult/Pediatric: 30 to 290 mmHg
Neonate: 30 to 140 mmHg
MAP Adult/Pediatric: 20 to 260 mmHg
Neonate: 20 to 125 mmHg
Diastolic Adult/Pediatric: 10 to 220 mmHg
Neonate: 10 to 110 mmHg
Accuracy According to AAMI SP10-2002 4.4.5.2
B, accuracy of NIBP parameter was
validated against the intra-arterial
method
1
.
Default initial inflation
pressure
Adult/Pediatric: 135 ± 15 mmHg
Neonate: 100 ± 15 mmHg
Over pressure allowed by
independent safety
controller
Adult/Pediatric: 300 ± 6 to 330 mmHg
Neonate: 150 ± 3 to 165 mmHg
1. Blood pressure measurements determined with this device are
equivalent to those obtained by an intra-arterial blood pressure
measurement device, within the limits prescribed by the American National
Standard, Manual, electronic, or automated sphygmomanometers