60
Cautions
PACKAGING DISPOSAL - Dispose of packaging material,
observing applicable waste control regulations.
Do not use or store equipment outside the specified
temperature, humidity, or altitude ranges.
DISPOSAL - At the end of its service life, the product described in
this manual, as well as its accessories, must be disposed of in
compliance with the guidelines regulating the disposal of each
product. If you have any questions concerning disposal of a
product, please contact GE or its representatives.
Cleaning schedules
NOTE: For details about the cleaning, disinfecting, and sterilizing the
accessories, see the instructions for use accompanying the
accessory.
Daily and between patient
Wipe the monitor and module surfaces with permitted cleaning
agents.
Wipe the ECG trunk cable, NIBP cuff and cables, and SpO
2
sensors with permitted cleaning agents. Avoid excessive use of
liquids.
Change or sterilize all airway and other invasive patient
accessories.
Clean, disinfect, or sterilize reusable temperature probes.
Empty the water trap.
Use a new Entropy sensor for each patient.
Check that the accessories, cables, and monitor are clean and
intact.
Check that the monitor battery is charged.
Every two months
Change the water trap when the ‘Replace D-Fend’ message
displays.
Perform gas calibration when gas module is in continuous using,
see Starting calibration.
Every six months
Condition the batteries, see Conditioning the batteries.
Perform gas calibration, see Starting calibration.