61
Permitted cleaning agents
The exterior surface can be cleaned with the following disinfecting
and sterilizing agents:
•Tap water
Distilled water
•Ethyl alcohol 99.5%
Ethyl alcohol 95 to 96%
Ethyl alcohol 90% with methyl alcohol 10%
•Ethyl alcohol 80%
Isopropyl alcohol 60%
•Phenol 2%
Purified benzine
•Chloramine 5%
Glutaraldehyde 2%
NOTE: The monitor does not directly expose the user or patient to
natural latex rubber and/or PVC (polyvinyl chloride).
General cleaning instructions
To clean the monitor, module, displays, and other parts, complete
the following procedure:
1. Turn off the monitor.
2. Disconnect the power cord.
3. Remove all cables and batteries (if applicable) and close battery
door(s).
4. Use a soft, lint-free cloth with one of permitted cleaning agents
to wipe the exterior surface.
5. Wipe off the cleaning solutions with a clean, lightly moistened
cloth.
6. Dry thoroughly with a dry, lint-free cloth and let air dry for at
least 30 minutes.
7. Reconnect the power cord.
8. Turn on the monitor.
NOTE: Check ventilation holes and clean if necessary.
NOTE: Any contact of cleaning agents with metal parts may cause
corrosion.
NOTE: Make sure not to leave any liquid spills on any metal part.
NOTE: Do not damage or bend connector pins when cleaning or
drying.
NOTE: For information regarding materials used in accessories, see
the instructions for use accompanying the accessory.