GE MEDICAL SYSTEMS CT9800 QUICK SYSTEM
Rev. 14 Direction 18000
4-3-62
Current Configuration Facts:
Logical Logical Logical Logical
Unit 0 Unit 1 Unit 2 Unit 3
Disk Type OP-Optical Disk OP-Optical Disk OP-Optical Disk OP-Optical Disk
Physical Units 0 1 2 3
Logical Heads 1 1 1 1
Sectors 128 128 128 128
Cylinders 14670 14670 14670 14670
Heads Offset 0 0 0 0
Usable Blocks 1877632 1877632 1877632 1877632
Platter ID GE service: John Doe
BMC Priority 4
Throttle burst rate 32
Break count 4
Base Platter number 1
Number of Optical Drives 1
Running in RDOS system YES
Enter Command (? to see choices): U<CR>
..EEPROM updated successfully.
Enter Command (? to see choices): Q<CR>
After the program has terminated, you must press IORESET to re-initialize the
controller with the new values.
Do you want to configure another ODSS controller (Y/[N])? N<CR>
Press Reset on the CPU front panel and reboot the tape.
The UTILITY Program
FILE # PROGRAM
2 GEMOS CONFIGURATOR
3 GEMOS UTILITY
FILE NUMBER? 3 <CR> Enter 3 to load the Utility Program.
A - DO ALL: FORMAT, INSTALL FIRMWARE, DKINIT
B - FORMAT MAGNETIC DISK See GUIDELINES FOR USAGE
C - INSTALL ARZ FIRMWARE ON MAGNETIC DISK See GUIDELINES FOR USAGE
D - INSTALL SCSI FIRMWARE ON MAGNETIC DISK See GUIDELINES FOR USAGE
E - INITIALIZE (DKINIT) THE MAGNETIC DISK See GUIDELINES FOR USAGE
I - INQUIRE THE DRIVE See GUIDELINES FOR USAGE
R - (RE) INITIALIZE THE CONTROLLER
P - PLATTER COMPLETION (PURGE OF MAGNETIC DISK) SEE GUIDELINES FOR USAGE
H - HELP
L - LOGGING TO THE PRINTER
Q - QUIT
CHOICE? A<CR> Enter A to format and install microcode on the MDD.
Note: This procedure requires approx. 15 Min. to complete.