GE MEDICAL SYSTEMS CT9800 QUICK SYSTEM
Rev. 14 Direction 18000
4-3-63
This function will erase the data currently on the magnetic disk. You should
only run it if you DO NOT have an active platter associated with the magnetic
disk. Any Platters which have been “completed” are not active.
Enter Y if you wish to proceed with this function. Y<CR>
Enter the device code for disk controller [34]: <CR>
Controller Init Routine
-- INQUIRY COMMAND
-- FORMATTING
-- INSTALLING ARZ-I Firmware
-- INSTALLING SCSI Firmware
-- INITIALIZING THE DISK
-- ALL FUNCTIONS COMPLETE
ENTER C TO CONTINUE: C<CR>
A - DO ALL: FORMAT, INSTALL FIRMWARE, DKINIT
B - FORMAT MAGNETIC DISK
C - INSTALL ARZ FIRMWARE ON MAGNETIC DISK
D - INSTALL SCSI FIRMWARE ON MAGNETIC DISK
E - INITIALIZE (DKINIT) THE MAGNETIC DISK
I - INQUIRE THE DRIVE
R - (RE) INITIALIZE THE CONTROLLER
P - PLATTER COMPLETION (PURGE OF MAGNETIC DISK)
H - HELP
L - LOGGING TO THE PRINTER
Q - QUIT
CHOICE? R<CR> You must reinitialize the controller after a format
Enter the drive code for disk controller [34]: <CR>
Controller Init Routine
--Controller reinitialized
ENTER C TO CONTINUE: C<CR>
Menu is reposted on CRT
CHOICE? I<CR> Check controller to MDD communication
--Inquiry command
Enter unit number:3<CR> For MDD
- Vendor ID is CDC
- Product ID is 94161-155
- Product revision level is XX
ENTER C TO CONTINUE: C<CR>
Menu is reposted on CRT
CHOICE? I<CR> Check controller to MDD communication
--Inquiry command
Enter unit number:0<CR> For ODD
Inquiry data not available Indicates successful communication
ENTER C TO CONTINUE: C<CR>
CHOICE? Q<CR> Exit program