GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 3 Direction 18000
8-6-13
1-3 DEADSTART EXAMPLES
The following examples illustrate how board failures effect the DEADSTARTS results. These examples were
obtained by removing the indicated board and running DEADSTARTS. They may prove useful if you encounter
similar failures when running Deadstarts. Deadstarts is a verify good service tool once you learn how to use it
properly for troubleshooting. You should read the preceding sections to learn more about the various sections of
the Deadstart tests.
DIR CTDS *** NO MASTER BOARD ***
R
UPAP
*** Deadstart Program (DS) *** Revision 1.05
Configuration Information
Processor Type = ST-X
01 Compute Head(s)
02 Back Projector board(s)
ACU Rounding Mode = ST-100
SMP MUX NOT PRESENT
Main Memory 1.0 Mega words
*E* Transfer to AP did not succeed. *E*
*E* Transfer to AP did not succeed. *E*
*E* Transfer to AP did not succeed. *E*
*E* Transfer to AP did not succeed. *E*
*E* Transfer to AP did not succeed. *E*
*E* RESET Failed *E*
*** Deadstarts Complete ***
etc.