3-5 Initial ZEBRA Disk Drive Check (Continued)
e. When the format portion of the program is done, the program will write and read patterns to the entire
pack looking for bad blocks. Each pass will take approximately 25 minutes, 50 minutes for Dual Drive.
During the course of the program, it is possible to monitor progress by typing L. The following is a typical
printout with no errors:
L
RUN START TIME WAS 8/6/82 13:44
DRIVE 0
SECTORS READ 64
SECTORS WRITTEN 586280
BAD SECTOR
CYL HEAD SECTOR
ADDRESS ERROR
CYL HEAD SECTOR
0 BAD SECTORS ON CYLINDER 0
0 BAD SECTORS MAX ON CYLINDER 0
0 BAD SECTORS MAX ON SURFACE 0
0 BAD SECTORS TOTAL
0 HARD DATA ERRORS > 11 BITS
Dual Drive systems, this printout will be followed immediately by the similar status for Drive 1.
If an ECC error is detected, a message similar to the following will be posted:
READ UNIT: 0
CYL 646 HEAD 5 SECT 16 #SECT 6
DIA STATUS = 060601 ECC DETECTED ERROR
ADDR WORD # GOOD BAD WORD
021273 000176 055555 015555
COUNT 1 HARD_
If program halts after an error printout, check switch register by typing ’M’ and ensure switch 1 is set to 1