DigitalFlow™ GS868 Service Manual 19
Chapter 2. Error Codes and Screen Messages
2.6 E4: Signal Quality
Problem: The signal quality is outside the limits programmed in the SETUP submenu of the User Program.
Cause: The peak of the upstream or downstream correlation signals has fallen below the correlation peak limit, as
set in the
SIGNL submenu. This may be caused by a flowcell or electrical problem.
Action: Check for sources of electrical interference and verify the integrity of the electronics console by
temporarily substituting a test flowcell that is known to be good. Check the transducers and relocate them,
if necessary. See Chapter 3, Diagnostics, for instructions.
2.7 E5: Amplitude Error
Problem: The signal amplitude exceeds the limits programmed in the SETUP submenu of the User Program.
Cause: Excessive levels of an attenuating gas, such as CO
2
, are present in the flowcell. Solid or liquid particulates
may be present in the flowcell.
Action: Refer to Chapter 3, Diagnostics, to correct any flowcell problems.
2.8 E6: Cycle Skip, Accel.
Problem: The acceleration exceeds the limits programmed in the SETUP submenu of the User Program.
Cause: This condition is usually caused by poor flow conditions or improper transducer alignment.
Action: Refer to Chapter 3, Diagnostics, to correct any flowcell and/or transducer problems.
2.9 E7: Analog Out Error
Problem: The current in the analog output circuit exceeds the limits for the analog output port.
Cause: The output load exceeds the specified limits for the analog output port.
Action: Make sure the output load is <550 ohms for the Slot 0 analog outputs or is <1000 ohms for an analog
outputs option card.
2.10 E8: Temp In
Problem: This message indicates a temperature input error.
Cause: The temperature exceeds the specified limits for the analog inputs option card or no input device is
connected.
Action: Check the temperature transmitter and the connecting cable. Refer to Chapter 1, Calibration, and
recalibrate the analog inputs on the option card.

Table of Contents

Other manuals for GE DigitalFlow GF868

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE DigitalFlow GF868 and is the answer not in the manual?

GE DigitalFlow GF868 Specifications

General IconGeneral
BrandGE
ModelDigitalFlow GF868
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals