DigitalFlow™ GF868 Service Manual (1 and 2-Channel) 19
Chapter 2. Error Codes and Screen Messages
E5: Amplitude Error
Problem: The signal amplitude exceeds the limits programmed in the SETUP submenu of the User Program.
Cause: Excessive levels of an attenuating gas, such as CO
2
, are present in the flowcell. Solid or liquid particulates 
may be present in the flowcell.
Action: Refer to Chapter 3, Diagnostics, to correct any flowcell problems.
E6: Cycle Skip, Accel.
Problem: The acceleration exceeds the limits programmed in the SETUP submenu of the User Program.
Cause: This condition is usually caused by poor flow conditions or improper transducer spacing.
Action: Refer to Chapter 3, Diagnostics, to correct any flowcell and/or transducer problems.
E7: Analog Out Error
Problem: The current in the analog output circuit exceeds the limits for the analog output port.
Cause: The output load exceeds the specified limits for the analog output port.
Action: Make sure the output load is <550 ohms for the Slot 0 analog outputs or is <1000 ohms for an analog 
outputs option card (slots 1-6).
E8: Temp In
Problem: This message indicates a temperature input error.
Cause: The temperature exceeds the specified limits for the analog inputs option card or no input device is 
connected.
Action: Check the temperature transmitter and the connecting cable. Refer to Chapter 1, Calibration, and 
recalibrate the analog inputs on the option card.
E9: Press In
Problem: This message indicates a pressure input error.
Cause: The pressure exceeds the specified limits for the analog inputs option card.
Action: Check the pressure transmitter and the connecting cable. Refer to Chapter 1, Calibration, and recalibrate 
the analog inputs on the option card.