EasyManua.ls Logo

GE DigitalFlow GF868 - Accessing the Calibration Menu; Calibrating the Low End of the Output Range; Calibrating the High End of the Output Range; Testing the Analog Output Linearity

GE DigitalFlow GF868
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 1. Calibration
4 DigitalFlow™ GF868 Service Manual (1 and 2-Channel)
1.3.2 Calibrating the Low End of the Output Range
1. Press [F1] to calibrate the low end of the output range.
2. Press [F1] or [F2] to adjust the ammeter reading UP or DOWN, until a 4 mA reading is achieved. If you press [F3],
Numer, enter a numeric value for the calibration point and press [ENT].
3. Press [F4] to STORE the setting, or press and [F1] to ABORT the calibration.
Note: If the ammeter reading cannot be adjusted within 5.0 A of the 4 mA setting, contact the factory for assistance.
1.3.3 Calibrating the High End of the Output Range
1. Press [F2] to calibrate the high end of the output range.
2. Press [F1] or [F2] to adjust the ammeter reading UP or DOWN, until a 20 mA reading is achieved. If you press [F3],
Numer, enter a numeric value for the calibration point and press [ENT].
Note: If the ammeter reading cannot be adjusted within 5.0 A of the 20 mA setting, contact the factory for
assistance.
3. Press
[F4] to STORE the setting, or press and [F1] to ABORT the calibration.
Note: Skip the next section if the linearity test will not be performed at this time. Proceed to Procedure Options on the
next page.

Table of Contents

Other manuals for GE DigitalFlow GF868

Related product manuals