EasyManuals Logo

GE DigitalFlow GF868 Service Manual

GE DigitalFlow GF868
74 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #14 background imageLoading...
Page #14 background image
Chapter 1. Calibration
4 DigitalFlow™ GS868 Service Manual
1.4 Accessing the Calibration Menu
1. Press the [CAL] key to enter the Calibration Program.
2. Press
[Fx] to calibrate the desired slot. (The option bar will include a slot listing for each installed option card.)
3. Press
[Fx] to select the desired output.
IMPORTANT: The procedure for calibrating all the outputs is the same. However, when calibrating a different output, be
sure to reconnect the ammeter to the appropriate pins on terminal block I/O. See Figure 1 on page 2 or
Figure 2 on page 3 for the correct pin numbers.
1.4.1 Calibrating the Low End of the Output Range
1. Press [F1] to calibrate the low end of the output range.
2. Press
[F1] or [F2] to adjust the ammeter reading UP or DOWN, until a 4 mA reading is achieved. If you press [F3],
Numer,
enter a numeric value for the calibration point and press [ENT].
3. Press [F4] to STORE the setting, or press and [F1] to ABORT the calibration.
Note: If the ammeter reading cannot be adjusted within 5.0 A of the 4 mA setting, contact the factory for assistance.
1.4.2 Calibrating the High End of the Output Range
1. Press [F2] to calibrate the high end of the output range.
2. Press
[F1] or [F2] to adjust the ammeter reading UP or DOWN, until a 20 mA reading is achieved. If you press [F3],
Numer
, enter a numeric value for the calibration point and press [ENT].
Note: If the ammeter reading cannot be adjusted within 5.0 A of the 20 mA setting, contact the factory for
assistance.
3. Press
[F4] to STORE the setting, or press and [F1] to ABORT the calibration.
Note: Skip the next section if the linearity test will not be performed at this time. Go to Procedure Options on page 5.
1.4.3 Testing the Analog Output Linearity
Note: If you exited this menu and are returning, follow steps 1 through 3 in Accessing the Calibration Menu on page 4.
1. Press
[F3] to TEST the linearity of the currently selected analog output.
2. Check the ammeter reading at the 50% output level.
3. Then, enter a different output level (0-100%) and press
[ENT].
4. Check the ammeter reading at this setting and press
[ENT] when done.
Table 1 on page 5 lists the expected ammeter readings at various % Full Scale settings, for both 4-20 mA and 0-20 mA
scales. Refer to this table to verify the accuracy of the ammeter readings taken above.

Table of Contents

Other manuals for GE DigitalFlow GF868

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE DigitalFlow GF868 and is the answer not in the manual?

GE DigitalFlow GF868 Specifications

General IconGeneral
BrandGE
ModelDigitalFlow GF868
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals