K0430 Issue No. 2 8
Change Pressure Units
To change the pressure units proceed as follows. If the four units displayed are not the
units required, press TASK and select any task, other than BASIC, press SETUP and
proceed as detailed on page 36. To return to BASIC mode, press TASK and select BASIC.
In BASIC mode, the unit is configured to carry out basic Pressure to Voltage (P to V) or
Pressure to Current (P to I) tests, a typical test procedure follows:
Voltage and Current Measurements
Connect the electrical input sockets as follows for voltage and current measurements.
Use the test leads provided and DO NOT push bare wires into the sockets.
55mA
Vin
mA in
mA Sink
max 30V
max
+
-
V
55mA
Vin
mA in
mA Sink
max 30V
max
+
-
Current
Maximum input current = 55mA dc (at 30 V dc)
Voltage
Maximum applied voltage = 30V dc.