K0430 Issue No. 2 26
mA Ramp
Press the OUTPUT key and select mA Ramp as shown previously in mA Step.
Define ramp required by entering START and END current values as shown below:
Connect an external power source as shown on page 24.
•Press START (F1) to run the program. A status display CHECK LOOP indicates a
fault in the external loop i.e. supply fault or open circuit.
Note: The ramp cycle (min to max or max to min), is approximately 60 seconds.
•Press STOP (F1) when running to stop at any point. Press CONTINUE (F1) to
resume from point of pause or RESET (F2) to return to start point.