K0430 Issue No. 2 30
Data Log Operations
Data Log is a special application of store mode which enables the calibrator to either
automatically log displays at preset time intervals or to manually log a display on
operation of the STORE key. Logged data is written to a user specified file.
To set-up a Data Log file, proceed as follows.
Select a task, other than BASIC. If using ADVANCED, set-up the required output
parameters.
•Use SETUP to select Data Log from the Store Mode Menu (See page 36).
Auto Log (Timer)
Press STORE and set-up the Data Log file parameters as shown below. Use CHANGE
VALUE (F1) followed by cursor keys to set field values. For Auto Log, set-up TRIGGER field
to PERIODIC.
Manual Logging
Enter the file details as shown above and select KEYPRESS for TRIGGER field. Screen
reverts to displayed parameters showing set-up file as shown below:
Use a combination of STORE and LOG NOW (F1) to log events as follows.
Memory Operations Datalog