MA-028 Rev. 2.1, 11-Jun-20
Moisture (level and average) and analog High contacts
Moisture (level and average) and analog High-High contacts
Relay contact maximum switching capacity:
Resistive load: 50 VA, 60 W (power factor = 1)
Relay contact maximum operating current:
Resistive load: 3 A (ac/dc)
Communications
Ports
One RS-232 port for local laptop computer connection
One RS-485 port for connection to local Hydran network or
remote communication
Optional
Communications
Port
1 of the following to be specified when ordering:
Integrated serial to Ethernet 10/100BaseT; Part Number
17046
Integrated Fiber Optic Ethernet 100BaseFX, ST Multimode;
Part Number 18321
Integrated dial-up Modem V.92/56k; Part Number 16868
Optional I/O
Up to 4 of the following to be specified when ordering:
General-purpose 420 mA analog input; 2,000 V RMS
isolation; Part Number 16528
Analog 420 mA output; 500- maximum loading; 2,000 V
RMS isolation; Part Number 16529
Dual digital inputs for dry contacts; internal wetting 24 VDC;
2,000 V RMS isolation; Part Number 17464
MISCELLANEOUS
Enclosure
Type NEMA 4X
Cast aluminum, powder pint, white
External dimensions: 315 mm long x 219 mm wide x 196 mm
high (12.4 x 8.63 x 7.72 in)
Electronic
Modules
Totally encased CPU and I/O electronics
Sensor Heating
Heater plate to maintain sensor between 15 and 65 °C (59 and
149 °F) in normal transformer operating conditions to force oil
convection
Oil Temperature
Oil at the valve: -40 to +90 °C (-40 to +194 °F)
Oil at the valve up to 105 °C (221 °F), with optional finned,
high-temperature adaptor, Part Number 18065

Other manuals for GE Hydran M2-X

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Hydran M2-X and is the answer not in the manual?

GE Hydran M2-X Specifications

General IconGeneral
BrandGE
ModelHydran M2-X
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals