MA-041 ā DGA 900 Plus - Installation & Commissioning Manual - Rev 1.1 4-Mar-22 Page 41 of 99
5. Insert the coaxial cable through the
connector fitting of the Bushing
Adaptor cover.
6. Connect the wires to connector J3 of
the Bushing Adaptor board.
7. J3 is a spring-cage type connector.
Open the terminal point with a small
flat insulated screwdriver.
8. The central white wire must be
inserted into the inner terminal.
9. The braid shield must be inserted into
the outer terminal.
10. A fully connected Bushing Adaptor is
shown.
At the cabinet end, repeat the steps 1-4 above and use the same pins to terminate the
cable.
9.4.1.1.5 Connecting to the Grounding Switch
To connect to the grounding switch, follow the steps as outlined in Table 9-5.
Table 9-5: Connecting to the grounding switch
The central white wire must be connected
to the core (CO) connector.
The braid shield must be connected to the