Page #97 background imageLoading...
Page #97 background image
GE
D
IRECTION 5273008-100, REVISION 9 LOGIQ C SERIES BASIC SERVICE MANUAL
Chapter 4 Functional Checks 4 - 29
4-3-14-4 Archiving Images
1.) Insert the archive media. To format the archive media, select the Utility button on the
Keyboard.
2.) Select Connectivity, then Removable Media.
3.) Format the CD. Verify the format if desired.
4.) Images will be moved from the hard drive by date. Therefore, the best way is to label media by
date.
NOTE: Images will be moved from the hard drive by date. Therefore, the best way to label media is by
date. When images are moved to the archive media, they will be deleted from the system hard
drive. However, the patient database (backed up earlier) maintains pointers to the location of
the images on the archive media.
5.) Select Backup/Restore, then EZBackup/Move.
6.) Select “Backup File Older Than in Days”.
Figure 4-20 Format CD Screen
Figure 4-21 EZBackup/Move
Choose the days
which want to older
than current exam
date.