GE HEALTHCARE
DIRECTION 5162630, REVISION 3 LOGIQ™ S6 SERVICE MANUAL
Section 5-5 - Common Service Platform 5-31
Temperature
Error (!!)
The error occurred in the temperature with “!!” mark. In the time when the error occurred, the
message “Temp is above safe limit. Check air-filter! Shutdown in #seconds.” appears in the status
bar and the system is automatically shut down.
The temperature which the error occurs:
• 75 degrees or higher for Rack Temperature (1)
• 65 degrees or higher for Rack Temperature (2)
• 65 degrees or higher for HV Temperature
Warning (@@)
The warning occurred in the temperature with “@@” mark. In the time when the warning occurred,
the message “This system is overheated! Check air-filter!” appears in the status bar.
The temperature which the warning occurs:
• 70 degrees or higher for Rack Temperature (1)
• 60 degrees or higher for Rack Temperature (2)
• 60 degrees or higher for HV Temperature
Info (^^)
The information occurred in the temperature with “^^” mark to call attention. In the time when the
information occurred, the message “Approaching the overheat limit! Check air-filter!” appears in the
status bar.
The temperature which the information occurs:
• 65 degrees or higher for Rack Temperature (1)
• 50 degrees or higher for Rack Temperature (2)
• 50 degrees or higher for HV Temperature
Others Temperature information
Probe Displays the probe name and connector #.
Board Displays the Part #, Board #, and Dip switch revision.
DICOM
Whenever DICOM data are sent, logs are reported. An error log is reported in an error status and an information log is
reported in a normal status. (The log is command or status information specified by DICOM statement.)
Table 5-47 Error Log Descriptions
Logs Descriptions