DIRECTION 5162630, REVISION 3 LOGIQ™ S6 SERVICE MANUAL
Section 6-3 - Reloading the Probe Data 6-7
Section 6-3
Reloading the Probe Data
This procedure will be used for troubleshooting when a image sensitivity is NOT recovered. For details,
refer to Section 7, Troubleshooting.
This rewrites the probe data (probe delay data for beam forming) saved on the flash memories of DDBF
64 (DDBF128) and BTRAP boards.
1.) Insert the Service Dongle into the USB port, located at the rear panel of the scanner.
2.) Power ON the scanner. Verify that the scanner boots up with no error.
3.) Tap ON/OFF the power switch.
4.) Click on Exit.
Figure 6-80 Inserting the Service Dongle
Figure 6-81 System-Exit Screen
USB port