DIRECTION 5162630, REVISION 3 LOGIQ™ S6 SERVICE MANUAL
Section 6-3 - Reloading the Probe Data 6-9
Section 6-3 Reloading the Probe Data (cont’d)
8.) Click on Exit to window.
9.) The Windows Desktop screen is displayed.
10.)Erase the file, “D:\Pegasus\target\resources\Pegasus\ProbeInfo.res.”
11.)Erase the above file completely from Recycle bin, then restart the scanner.
Figure 6-84 Exit to Window
Figure 6-85 Windws Desktop