GE HEALTHCARE
DIRECTION 5162630, REVISION 3 LOGIQ™ S6 SERVICE MANUAL
Section 6-4 - Monitor and LCD Adjustments 6-13
6-4-2-1 Adjusting Parameters (cont’d)
Adjust the Contrast value
6.) Adjust the Contrast value by pressing button (2) and (3).
Exit the Contrast Menu
7.) Press the button (1) to go back to the Color menu.
8.) Press the button (3) 5 times to select Return at the lower left end.
9.) Press the button (1) to go back to the Advanced Menu.
10.)Press the button (3) 4 times to select Exit at the lower left.
11.)Press the button (1) to exit.
6-4-2-1-3 Color Temperature
Range: 9000K / 11000K / 13000K / 15000K
Property: Reddish <-> Bluish.
Factory default: 11000K
Adjustment: Change the value in Advanced Menu according to the procedure below.
Go into the Color Temperature Menu
1.) Press and hold the button (1) for 15 seconds, and the Advanced Menu appears.
2.) Press the button (3) once to select the Color menu ([RGB] icon).
3.) Press the button (1) to go into the Color menu. The upper left end icon is highlighted in
the menu.
4.) Press the button (3) once to select Temperature menu ([K] icon).
5.) Press the button (1) to go into the Temperature menu.
Color Temperature value adjustment
6.) Adjust the Color Temperature value by pressing the button (2) and (3).
Exit the Color Temperature Menu
7.) Press the button (1) to go back to the Color menu.
8.) Press the button (3) 4 times to select Return at the lower left end.
9.) Press the button (1) to go back to the Advanced Menu.
10.)Press the button (3) 4 times to select Exit at the lower left end.
11.)Press the button (1) to exit.