GE HEALTHCARE
DIRECTION 5162630, REVISION 3 LOGIQ™ S6 SERVICE MANUAL
8-10 Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
8-3-5-4 Saving Connectivity
Before starting upgrading, write down
the Connectivity Setting for back-up.
1.) Touch Utility > Connectivity on the touch panel and click the TCP/IP tab.
2.) Write down the following parameters:
Figure 8-117 TCP/IP Screen
Table 8-63 Saving Connectivity
Parameters Descriptions (Entry for Stand-Alone System)
Computer Name
Enable DHCP This parameter must NOT be selected.
IP-Address (192.168.1.2)
Subnet Mask (255.255.255.0)
Default Gateway (192.168.1.1)
Note : For a stand-alone system, preset values of IP adress, subnet mask, and default gateway, shown above,
must entered.
Remove the
Check mark.