GE HEALTHCARE
DIRECTION 5162630, REVISION 3 LOGIQ™ S6 SERVICE MANUAL
Section 10-3 - Periodic Maintenance Schedule 10-3
10-3-1How often should PMs be performed? (cont’d)
Table 10-2 Periodic Maintenance Schedule
Service at Indicated Time Daily Weekly Monthly
Per
Facilities
QA
Program
Notes
Clean Probes •* * or before each use
Clean Probe Holders
Clean Air Filter
more frequently depending on
your environment
Inspect AC Mains Cable Mobile Unit Check Weekly
Inspect Cables and Connectors
Clean Console
Clean Monitor and Touch Panel
Inspect Wheels, Casters, brakes and
Swivel Locks
Mobile Unit Check Daily
Check Control Panel Movement
Mobile Unit Check Daily
Console Leakage Current Checks
also after corrective
maintenance or as required by
your facilities QA program
Peripheral Leakage Current Checks
Surface Probe Leakage Current Checks
Endocavity Probe Leakage
Current Checks
Transesphongeal Probe Leakage
Current Checks
Surgical Probe Leakage
Current Checks
Measurement Accuracy Checks
Probe/Phantom Checks

Table of Contents

Other manuals for GE LOGIQ S6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S6 and is the answer not in the manual?

GE LOGIQ S6 Specifications

General IconGeneral
BrandGE
ModelLOGIQ S6
CategoryMedical Equipment
LanguageEnglish

Related product manuals