GE HEALTHCARE
DIRECTION 5162630, REVISION 3 LOGIQ™ S6 SERVICE MANUAL
Section 10-3 - Periodic Maintenance Schedule 10-3
10-3-1How often should PMs be performed? (cont’d)
Table 10-2 Periodic Maintenance Schedule
Service at Indicated Time Daily Weekly Monthly
Per
Facilities
QA
Program
Notes
Clean Probes •* * or before each use
Clean Probe Holders
•
Clean Air Filter •
more frequently depending on
your environment
Inspect AC Mains Cable • Mobile Unit Check Weekly
Inspect Cables and Connectors •
Clean Console •
Clean Monitor and Touch Panel •
Inspect Wheels, Casters, brakes and
Swivel Locks
• Mobile Unit Check Daily
Check Control Panel Movement
• Mobile Unit Check Daily
Console Leakage Current Checks •
also after corrective
maintenance or as required by
your facilities QA program
Peripheral Leakage Current Checks •
Surface Probe Leakage Current Checks •
Endocavity Probe Leakage
Current Checks
Transesphongeal Probe Leakage
Current Checks
Surgical Probe Leakage
Current Checks
Measurement Accuracy Checks •
Probe/Phantom Checks •