GE LOGIQ V5/LOGIQ V3
D
IRECTION 5496012-100, REVISION 3 BASIC SERVICE MANUAL
Chapter 4 - Functional Checks 4 - 21
CF/PDI Auto 
Sample Volume
Yes Set the default value at Utility -> Imaging -> CF Mode.
CF/PDI Focus 
Depth
Yes
CF/PDI Frequency Yes
CF/PDI Auto 
Frequency
Yes
CF/PDI Center 
Depth
Yes
PDI Yes Power Doppler Imaging (PDI) is a color flow mapping 
technique used to map the strength of the Doppler signal 
coming from the tissue rather than the frequency shift of 
the signal. Using this technique, the ultrasound system 
plots color flow based on the number of reflectors that are 
moving, regardless of their velocity. PDI does not map 
velocity, therefore it is not subject to aliasing. 
TVI (Option) Yes Tissue Velocity Imaging (TVI) calculates and color-codes 
the velocities in tissue. The tissue velocity information is 
acquired by sampling of tissue Doppler velocity values at 
discrete points.The information is stored in a combined 
format with gray scale imaging during one or several 
cardiac cycles with high temporal resolution.
TVD Yes TVD: Tissue Velocity Doppler: basing on TVI mode, 
activate a sample volume of PW ventricular wall to get the 
spectral information of the sample section.
Table 4-5    Color Flow Mode Controls
Control Possible Bioeffect Description/Benefit