10 - 8 Section 10-4 - Tools Required
10-4-6 Outlet Test -Wiring Arrangement - USA & Canada
Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet 
tester and noting the combination of lights that are illuminated. Any problems found should be reported 
to the hospital immediately and the receptacle should not be used.
NOTE: No outlet tester can detect the condition where the Neutral (grounded supply) conductor and 
the Grounding (protective earth) conductor are reversed. If later tests indicate high leakage 
currents, this should be suspected as a possible cause and the outlet wiring should be visually 
inspected.
Figure 10-1  Typical Outlet Tester