IRECTION 5496012-100, REVISION 3 BASIC SERVICE MANUAL
4 - 26 Section 4-3 - General Procedure
4-3-13-2 Backup System Presets and Configurations
NOTE: Always backup any preset configurations before a software reload. This ensures that if the presets need
to be reloaded, after the software update, they will be the same ones the customer was using prior to
service.
1.) Insert a formatted media into the drive.
2.) Enter enter Utility-> System-> Backup/Restore.
NOTE: If you are not logged in as GE Service or with administrator privileges, the Operator Login
window is displayed. Log on with administrator privileges.
3.) In the Backup list, select Patient Archive, Report Archive, User Defined Configuration and
Service.
4.) In the Media field, select CD/DVD (or USB memory device).
5.) Select Backup, as shown in Figure 4-18 on page 4-26.
The system performs the backup. As it proceeds, status information is displayed on the Backup/Restore
screen.
Figure 4-18 Backup/Restore Menu

Table of Contents