Marquette Hellige GmbH MAC 1100/1200 V 1.1 Page 25
227 492 20 D - 0002
MO_SU/
Pin
Number
Signal input/output Function Definition
1 +UVERS Out
p
ut Su
pp
l
y
Volta
g
e + 26
,
5V
2 EN_MOSU Output Enable Modem Supply Open, high: enable
Low: disable
3 GND24V Out
p
ut Ground
Connector denotation: RS232_E/
Type: male connector, 2x 5-pin, 180 °
The function of the individual pins is given in the following table . The definition as an
input/output is seen with reference to PCB Control CS_CI.
RS232_E/
Pin
Number
Signal input/output Function Definition
1+5V Out
p
ut Lo
g
ic Su
pp
l
y
Reserved
2+5V Out
p
ut Lo
g
ic Su
pp
l
y
Reserved
3 GND Ground
4 GND Ground
5SMRXD1 In
p
ut Receive
(
RXD
)
Reserved
6SMTXD1 Out
p
ut Transmit
(
TXD
)
Reserved
7 GND Ground
8
V
MODEM In
p
ut Modem Su
pp
l
y
+8
V
9 EKGTRIG E Out
p
ut ECG Tri
gg
er Reserved
10 FERNST E In
p
ut Remote Start In
p
ut Reserved
2.6 Interfaces to peripherals
The MAC1200 has only 3 interfaces for peripherals:
Mains input
Patient input
RS-232 interface
The mains input interface on the device is a 3-pin standard cold appliance socket
connection, which is integrated into the mains input module. Connection to the mains
is effected via a 3-pin power cord with a non-fused earth conductor.
The mains input is a wide range input from 95Vac ... 240Vac. The mains input
module contains the two primary fuses.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE marquette MAC 1100 and is the answer not in the manual?

GE marquette MAC 1100 Specifications

General IconGeneral
BrandGE
Modelmarquette MAC 1100
CategoryMedical Equipment
LanguageEnglish

Related product manuals