Marquette Hellige GmbH MAC 1100/1200 V 1.1 Page 45
227 492 20 D - 0002
5 Troubleshooting tips
For all the following points the safety instructions in Section 4.1. have to be
observed!
Device cannot be switched on even though power plug is plugged in
- green power lamp LED is off and device cannot be switched on:
- Power cable defective or not plugged in correctly?
- Primary fuses in the mains input module defective?
- Connector from the mains input module to the mains switching power supply
plugged in correctly, or cable defective?
- Connector from the mains switching power supply to connector POSUP/ on the
PCB Control plugged in correctly, or cable defective?
- Keypad via connector KEYB/ on PCB Control plugged in correctly?
- 26,5V on the connector POSUP?
if no: - mains switching power supply defective
if yes: - PCB Control defective
- or keypad defective
- green power lamp LED is on, but device still cannot be switched on:
- PCB Control defective
- or keypad defective
Device cannot be switched on when being battery-operated only
- Is the battery depleted?
- Plug in power plug, green power lamp LED should glow and the device can be
activated,
if not: - refer to "Device cannot be switched on even though power
plug is plugged in”.
- By connecting the power plug, rapid charging is activated, LED1 on the PCB
Battery Charge is activated
if not: - PCB Battery Charge is defective
- With power plug plugged in, let unit charge for 10 minutes, then disconnect
power plug. Can the device be switched on and can a recording be started?
if yes: - Function OK., continue loading with plugged mains plug
if not: - Battery disconnected or cable defective?
- Battery defective (no capacity)?
- PCB Battery Charge defective ( Si1?)?