Marquette Hellige GmbH MAC 1100/1200 V 1.1 Page 27
227 492 20 D - 0002
2.6.1.2 Patient input
The connection to the patient cable is realized with a 15-pin female sub-D
Connector, implemented on the PCB Control CS_CI directly. The mechanical plastic
housing makes sure that for MAC1100/1200 only released patient cables can be
connected.
Connector denotation: EKG/
Type: 15-pin sub-D, female
EKG/
Pin
Number
Signal input/output Function Definition
1C2 In
ut chest lead V2
2C3 In
ut chest lead V3
3C4 In
ut chest lead V4
4C5 In
ut chest lead V5
5C6 In
ut chest lead V6
6 Shieldin
Out
ut shieldin
AVSS
7 Cable Input Identification: patient cable
used
“0“: patient cable used
8 Cable identification Input 5/10 -pin or 12 -pin cable
used
“0“: 12-pin cable
“1“: 5- or. 10-pin cable.
9R
RA
In
ut electrode ri
ht arm
10 L
LA
In
ut electrode left arm
11 F
LL
In
ut electrode left le
12 C1 In
ut chest lead V2
13 NST In
ut Nehb electrode NST
14 N
RF
Out
ut electrode ri
ht le
15 NAX In
ut Nehb electrode NAX