Marquette Hellige GmbH MAC 1100/1200 V 1.1 Page 35
227 492 20 D - 0002
3.10 Time and date
This section deals with the quick checking and setting of the time and date.
Pressing the “7" key calls up the following display:
Time (hh:mm) [18:25:29]
and
Date (dd.mm.yyyy) [03.03.1999]
X = End
To set time or date, select the corresponding time or date field by pressing the return
key. During entering the numbers, a single cursor is not displayed.
3.11 Electrode test
The electrode test is started by pressing the “8" key.
The following menu is displayed:
Electrode test
Terminate with any key
Select pace electrode (key ‘P’)
Pace electrode LA
Cable = 10 electrodes (or no cable)
The status of each electrode is displayed. Electrode OK or disconnected is displayed
as follows:
Status = 0 -> Electrode OK
Status = 1 -> Electrode disconnected
With the “P” key, the pace electrode can be selected. The selected pace electrode is
displayed.
The status “cable” indicates if a cable is connected to the patient ECG input or not.