Marquette Hellige GmbH MAC 1100/1200 V 1.1 Page 47
227 492 20 D - 0002
Paper transport functions, no printout
- Paper feed flap correctly engaged on both sides?
- All connectors for printhead plugged in?
- Operation with strongly depleted battery?
if yes: - plug in power plug
if no: - PCB Control defective
- or printhead defective
MAC 1200 only prints on the upper or lower section of the printout
- Paper feed flap is only engaged on one side.
MAC 1200 prints, but only baselines are printed out
- Electrodes applied correctly?
- Electrode cables plugged into the patient trunk cable terminal-box correctly?
- Patient trunk cable defective ( e.g., RL defective)?
- Contact problems at the patient input connector of the MAC1200?
if not: PCB Control defective.