Marquette Hellige GmbH MAC 1100/1200 V 1.1 Page 65
227 492 20 D - 0002
Signal inputs
isolated patient signal input, IEC type CF, high-voltage protection for all lead connections
and neutral electrode, interference compensation via neutral electrode, monitoring for open
leads
electrode connections for RA, LA, LL, LA, Vl to V6
input impedance for differential signals between any two electrode connections > 10 M
at 10 Hz
input impedance for common-mode signals referred to neutral electrode > 50 M up to
60 Hz
dynamic range for differential signals between any two electrode connections for AC
voltage ±10 mV, for superimposed DC voltage (polarization voltage) ±600 mV
dynamic range for common-mode signals referred to neutral electrode ±l V, referred to
chassis 263 V AC (rms.)
quiescent input current via any electrode connection for l k termination referred to
neutral electrode < 50 nA
patient leakage current (rms values) according to IEC, class CF: in normal condition
< 10 µA, in single-fault condition (e.g. patient in contact with line voltage) < 20 µA
non-destructive range for lead-electrode connections and the neutral electrode
connection referred to neutral electrode ±50 V, referred to chassis ±1500 V
pulse voltage resistance of all lead electrode connections and of the neutral electrode
connection referred to chassis (either polarity, e.g. defibrillation) 5000 V
monitoring of each electrode for open leads: RA, LA, LL, RL, Vl, V2, V3, V4, V5, V6
audio signal at printer start
Data interface
one serial RS232 interface for exchange of data with suitable external devices and software
handshake
RS232 interface (standard V.24 interface):
input voltage range. ± 15V max.
output voltage range ±5 V min.
interface protected from electrostatic discharge for ±10 kV max.