After The Exam Is Over
3-38 Versana EssentialUser Guide
5759238-100 English
Rev. 2
Troubleshooting
Please refer to Service Manual if other messages appear on the
monitor display.
Table 3-4: Error messages
The system has detected the lower air filter requires
cleaning. Please clean the lower air filter.
1. Shutdown the system. Disconnected the power cord.
Then clear the air filter according to ‘Cleaning the air
filters’ on page 3-35.
System temperature is too high. System will shut down.
1. Shutdown the system.
2. Clean the air filter according to ‘Cleaning the air filters’
on page 3-35.
System voltage fault. System will shut down.
1. Select OK and reboot the system.
2. If the same message appears after reboot, shut down
the system and turn off the breaker. Then turn on the
system according to ‘Power On’ on page 1-36.
System Error. Please reboot the system.
1. Select OK and reboot the system.
2. If the same message appears after reboot, shut down
the system and turn off the breaker. Then turn on the
system according to ‘Power On’ on page 1-36.