Probe Overview
Versana EssentialUser Guide 5-5
5759238-100 English Rev. 2
Probe Usage
For details on connecting, activating, deactivating,
disconnecting, transporting and storing the probes, See ‘Probes’
on page 1-40 for more information.
Care and Maintenance
Inspecting probes
Before each use
1. Inspect the probe's lens, cable, casing, and connector for
cracks, cuts, tears, and other signs of physical damage.
2. Test the functionality of the probe.
Figure 5-3. Probe parts
After each use
1. Inspect the probe's lens, cable, casing, and connector for
cracks, cuts, tears, and other signs of physical damage.
2. Look for any damage that would allow liquid to enter the
probe.
CAUTION
If any damage is found, DO NOT use the probe until it has
been inspected and released for further use by a GE service
representative.
1. Housing
2. Strain relief
3. Seal
4. Lens