Environmental Requirements
The system should be operated, stored, or transported within
the parameters outlined below. Either its operational
environment must be constantly maintained or the unit must be
turned off.
NOTE: You may get an overheating message with regard to fan speed.
Ensure adequate system/room ventilation.
Table 1-1: System Environmental Requirements
Operational (with
probe)
Storage (Versana
Essential)
Transport (Versana
Essential)
Temperature 10 - 40° C
50 - 104° F
-5 - 50° C
23 - 122° F
-5 - 50° C
23 - 122° F
Humidity 30 - 80% non-condensing 10 - 90% non-condensing 10 - 90% non-condensing
Pressure 700 - 1060hPa 700 - 1060hPa 700 - 1060hPa
The system cannot be used in OXYGEN rich environment.
Ensure that the probe face temperature does not exceed the
normal operation temperature range.